SBG01 Water Cooled Heat Flux Sensor

SBG01 Schmidt-Boelter/Gardon gauge. Heat flux sensor

The SBG01 is a water cooled high-intensity heat flux sensor (according to Schmidt-Boelter), intended for fire/flame measurement applications (e.g. flammability and smoke chamber testing). The SBG01 is compliant with ASTM E 622-83, ISO 5658, 5560 and 17554 measurement standards.

Add to Request

Applications
• Solar Trough & Solar Concentrated Power Tower Studies
• Flammability & Smoke Chamber Testing
• Curing & Smelting Furnace Applications

Engineered for fire and flame heat flux measurement research, the SBG01 is intended for high-intensity flux measurement applications. The SBG01 is suitable for use in applications dominated by thermal radiation. For applications environments with significant convection effects, additional considerations should be taken.

FEATURES
•Hardened water supply lines for improved durability
•Rugged stainless steel housing construction
•Scratch resistant detector coating (slightly lowered surface)

Employing entirely passive thermopile-based sensing technology, the SBG01 generates an output voltage signal that is proportional to the incoming heat flux received at the sensor detector level. The sensor housing is water cooled, providing a stable operational case temperature under high-intensity flux conditions up to 200 kW/m2.

SBG01 Specifications
ISO Classification: Second Class
Temperature range, cooling water: 10 to + 30°C
Cooling water flow: > 10 L/hr (normal tap water) 30 L p/hr @ 3 bar recommended
Working ranges kW/m2: 5,10,20,50,100,200
Response times
Working range 5, 10 kW/m2: < 450 ms (63%)
Working range 20, 50 kW/m2: < 250 ms (63%)
Working range 100, 200 kW/m2: < 200 ms (63%)
Maximum range: 150% of working range
Output signal: > 5 mV at working range
Spectral range: to 50µm
Field of view: 180 degrees
Emissivity: > 0.95
Calibration traceability: NIST

The SBG01 range consists of six (6) different SBG01 models; model selection depends on required sensitivity, response time and expected maximum intensity range (5, 10, 20, 50, 100 and 200 kW/m2 maximum). Compared to traditional Gardon and Schmidt-Boelter gauges, the SBG01 offers several performance advantages.

Ordering Information
SBG01 Water Cooled Heat Flux Sensor (please specify range and cable length)

sbg01-spec-sheet-2011

Weight 5 kg