A change with the EXO platform will be the BGA probes and Chlorophyll probes have been combined into a single sensor. There is no option to purchase one or the other.
This sensor is referred to as the “Total Algae” Sensor because having both chlorophyll and algae are important for estimating the total algae bio mass; the sensor will still output a BGA reading and a Chlorophyll reading independently.
SPECIFICATIONS
Blue-green Algae Phycocyanin
Range: 0 to 100 μg/L; 0 to 100 RFU;
Linearity: R2 >0.999 for serial dilution of Rhodamine WT solution from 0 to 100 μg/mL BGA-PC equivalents
Detection Limit: 0.04 μg/L PC
Response: T63<2 sec
Resolution: 0.01 μg/L; 0.01 RFU
Chlorophyll
Range: 0 to 400 μg/L Chl; 0 to 100 RFU
Linearity: R2 >0.999 for serial dilution of Rhodamine WT solution from 0 to 400 μg/L Chl a equivalents
Detection Limit: 0.09 μg/L Chl
Response: T63<2 sec
Resolution: 0.01 μg/L Chl; 0.01 RFU
Ordering Information:
599102-01 EXO Total Algae – PC Sensor, Ti “EXO Total Algae – PC Sensor, Ti
– Optimised for freshwater use – Phycocyanin
– Includes chlorophyll and blue green algae sensors in a single Sensor
– Incorporates wet-mate connector and sealed welded titanium design”