M1000 Optical Dissolved Oxygen Controller

InsiteIG Model 1000 Optical Dissolved Oxygen system is our pick for any mains powered DO system

The Model 1000 dissolved oxygen analyser is a unique system that combines advanced electronics with a solid-state optical sensor. No other dissolved oxygen system can compare with the features and benefits of this revolutionary design. The Model 10 sensor utilises an optical technique that does not consume oxygen like all standard membrane sensors

Add to Request

 

LEGACY MODEL- New Model available SCA 

The M1000 Dissolved Oxygen Analyser is a digital instrument designed for the continuous monitoring of dissolved oxygen in water and wastewater where parts per million accuracy is required. The instrument is designed to be used with the Model 10 sensor. The unit will display dissolved oxygen content in 0.01 ppm resolution over a range of 0.00 to 3.99 ppm and 0.1 ppm resolution over a range of 4.0 to 20.0 ppm. Temperature is displayed in 0.1 degree Celsius increments over a 0.0 to 50.0 degree Celsius range or 1 degree Fahrenheit increments over a 32 to 122 degree Fahrenheit range.The microprocessor based electronics of the Model 1000 analyser provide a high degree of flexibility and ease of use. Calibration is not required on a routine basis, nor is calibration required after initial start-up and commissioning.

Two isolated analogue outputs are standard. Three programmable set-point relays and one relay to control self-cleaning are also standard.

The sensor to be used with this analyser is an optical type sensor that measures the fluorescence and quenching reactions of a ruthenium complex that is immobilised in a sol-gel matrix.

SPECIFICATIONS
Power Requirements 115 VAC 60 Hz or 230 VAC 50 Hz
Accuracy 1% of reading or 0.02 ppm, whichever is greater
Sensitivity or Resolution 0.01 ppm
Stability (per 24 hr period) 0.01 ppm
Repeatability 0.01 ppm
Temperature Drift Not applicable
Zero Output Reading Not applicable
Consequence of Loss of Sample or Power User selectable
Measuring ranges 0.0 to 25.0 ppm
0 to 50 degrees C
Temperature compensation 0 to 50 degrees C
Calibration Method Not required nor recommend during initial start-up. Provisions for one-point in water calibration provided.

Ordering Information:
IIG-1000-10 Model 1000 analyser with Model 10 optical sensor
Four relays, dual isolated 4-20ma, and Modbus are standard outputs. The sensor comes with 25ft (7m) of cable. Custom cable lengths are available

Options
IIG-CABLE Additional sensor cable
IIG-AM1 Analyser mounting hardware
IIG-DO1 DO Sensor mounting hardware
IIG-IH1 Internal heater
IIG-PA Air Clean for plant supplied air
IIG- CA1 Compressor assembly for self-cleaning
IIG-1000-SENSOR D.O. Sensor only