STP01 ADVANTAGES
The STP01 has several advantages over existing designs:
• high accuracy gradient measurement by accurate positioning of the tc joints (+/- 1mm), and measurement of tc voltage relative to the junction at 50 cm (+/- 0.02 K is achievable)
• high accuracy and stability of the relative distance between sensors (+/- 0.5 mm)
• thin (0.6 mm only) construction, contrary to conventional stick designs, that doesn’t disturb the thermal flow pattern
• sensor to logger cabling is all copper; easily extendable.
• easy quality assurance & servicing: self-test is possible by heating and looking at the pulse response, saving servicing time.
EXPERIMENTAL USE: SOIL THERMAL CONDUCTIVITY MEASUREMENT
When using the heater, the tc pulse response can be used to calculate soil thermal conductivity at 3 depths; 5, 10 and 20 cm. The well established thermal needle or non-steady-state probe technique is applicable. For more theory on this method, see ask for the Hukseflux TP02 manual. (T = A.ln(t) +B, with T temperature, t time, A a function of geometry, power and thermal conductivity). The possibility to perform this measurement is an experimental option and the measurement accuracy is not specified by Hukseflux.
Suggested Use
• scientific grade measurement of soil temperature
• soil temperature gradient measurement
• energy balance & heat flux determination
SENSOR SPECIFICATIONS
Thermocouples: tc Cu-CuNi, type T
Temperature range: -30 to +70 °C
Depths below soil surface 2, 5, 10, 20 & 50 +/- 1 mm
Reference junction: Pt100 DIN class B
Thickness (nominal): 0.6 mm (2.5 at Pt100)
Required readout: 4 diff volt, 1 common ground, Pt100-4 or 3 wire
HEATER SPECIFICATIONS
Resistance (nominal): 200 Ω
Voltage input: 9-15 VDC
Duration of the self test: ± 10 min at 0.3 Watt
Average power consumption: 0.05 Watt
Required readout: 1 diff volt
Self-test recommendation once every 3 hours, at least every 24 hours
DATA ACQUISITION AND CONTROL
Solutions for measurement and control: See the STP01 manual or inquire at Hukseflux. Example programs for Campbell Scientific CR10X and CR1000 is available.
Ordering Information:
STP01Soil temperature profile
Options
Extended cable x meter (added to the standard length of 5 m)
Insertion tool IT01
See also model ST01