The meter has a sample and log mode, and will record an average daily value. Sample mode will record up to 99 manual measurements. Log mode will power the meter on/off to make a measurement every 30 seconds. Every 30 minutes the meter will average the sixty 30 second measurements and record the averaged value to memory. The meter can store up to 99 averages, once full it will start to overwrite the oldest measurement with new ones. An average daily value will be recorded from the 48 averaged measurements (making a 24 hr period). Sample and log measurements can be reviewed on the LCD display or by downloading the data to a computer, however, the average daily value can only be viewed by downloading the data to a computer. Downloading data to a computer requires the AC-100 communication cable (a standard USB cable will not work) and ApogeeAMS software.
Specifications
- Measurement Range 0 to 100 % Oâ‚‚
- Measurement Repeatability ± 0.1 % at 20.9 % O₂
- Non-linearity Less than 1 %
- Oxygen Consumption Rate 2.2 µmol O₂ per day at 20.9 % O₂ and 23 C
- Response Time 14 s (time required to read 90 % of saturated response)
- Operating Environment 0 to 50 C, less than 90 % non-condensing relative humidity up to 30 C, less than 70 % non-condensing relative humidity from 30 to 50 C, 60 to 140 kPa
- Meter Dimensions 12.6 cm length, 7.0 cm width, 2.4 cm height Sensor
- Dimensions 3.2 cm diameter, 6.8 cm length
- Mass 210 g Cable 2 m of two conductor, shielded, twisted-pair wire with santoprene rubber jacket (high water resistance, high UV stability, flexibility in cold conditions), additional cable available
- Warranty 4 years against defects in materials and workmanship
Ordering information
MO-200