Prolonged recording and reliability is ensured with premium quality EEPROM memory. 32k of memory allows up to 32,000 samples of data.
Our tough weld-sealed cases are rated IP56 and ensure a durable logger offering reliability in the harshest of environments and added security for sensitive information. A dust cover is supplied with the RH recorder for extended reliability of your RH sensors in harsh environments.
Temprecord software allows you to view both humidity and temperature recording on the one graph. Use the Temprecord Logger time and time again. The unit’s Lithium Thionyl Chloride battery provides a life expectancy of approximately 2 years.
SPECIFICATIONS
Memory size: 32k – records a total of 32000 samples per use. 16000 Temperature and 16000 Humidity recordings.
Sample period: Programmable between 2 seconds and 36 hours in 2 second increments
Start delay: Programmable between 2 seconds and 36 hours or use the start on date and time function
Start and Stop: Start / Stop button on logger or use the software and reader interface to start the logger
Recording duration (memory full): From 16 hours with a 2 second sample period set, to 2 years with a 36 hour sample period set
Temperature range: -40°C to +75°C
Temperature accuracy: +/-0.2°C over the range -20°C to +50°C, otherwise 0.5°C / 0.9°F
Humidity range: 0.00% to 100% RH
Humidity sensor accuracy: 3% RH between 12 to 85% RH, otherwise 5% RH between 0.00 to 12% and 85 to 100% RH
Inside range indicator: One flash from Green LED every ten seconds
Outside range indicator: One flash from Red LED every ten seconds
Display resolution: 0.01°C ~ 0.01°F
Battery type: Lithium Thionyl Chloride 3.6V
Dimensions; weight: 129 x 73.5 x 14mm; 83 grams
Case protection rating: Splash proof to IP56 (NEMA 4X). IP56 Protection Rating applies to the portion of the case that houses the electronics and battery, not the RH sensor itself. If the sensor has been exposed to water spray or jets of water you must dry the RH sensor using the procedure described in the Temprecord user manual to return the RH sensor back to factory calibration conditions.
Warranty Period : 12 months on logger from date of purchase ~ 2 years on battery
Ordering Information:
953170 RH HUMIDITY & TEMPERATURE Logger, 32k, Internal Sensors
953172 RH Logger with 1m Temperature Probe , 32K with YSI Sensor
952181 SR2 Reader Interface for PC (Serial RS-232 Connection)
953188 SR2 Reader Interface for PC (USB Port Connection