HD300 Thermo Anemometer with IR Temperature

Built-in IR Thermometer measures non-contact surface temperature

Built-in IR Thermometer Measures non-contact surface temperatures of objects unsafe or difficult to reach
Extech HD300 CFM Thermo-Anemometer. This handheld meter measures and displays air velocity, air flow (volume), temperature of air, and surface temperature (using the built-in non-contact IR thermometer). This meter is shipped fully tested and calibrated and, with proper use, will provide years of reliable service.

Add to Request

Features:

• InfraRed Thermometer measures remote surface temperatures to 932°F (500°C) with 30:1 distance to spot ratio and Laser pointer

• Simultaneous display of Air Flow in CFM/CMM or Air Velocity plus Ambient Temperature

• Easy to set Area dimensions are stored in the meters internal memory for the next power on

• Large (9999 count) LCD backlit display

• 20 point average for air flow

• USB port includes software

• Includes Windows® compatible software with cable, vane sensor with 3.9ft (120cm) cable, hard carrying case, and 9V battery

SPECIFICATIONS
Display: Dual Display Multi-function LCD with 9999 counts
Measurements Air Velocity: m/s, km/h, ft/min, knots, mph;
Air Flow: CMM (m3/min) and CFM (ft3/min);
Air Temperature (through vane) and Surface Temperature (using the IR thermometer function): °C and °F
Data Hold: Freezes displayed reading

Sampling rate: 1 reading per second
Sensors Air velocity/Air flow sensor: Conventional angled vane arms with low-friction ball bearing.
Air temperature sensor: Precision thermistor (built into the vane assembly); Surface temperature via non-contact IR sensor.

IR Distance to Spot ratio: 30:1
IR Spectral response: 6 to 14μm
IR Emissivity: 0.95 fixed

MIN-MAX-AVG Record and Recall: lowest, highest, and average readings
Auto Power OFF: Automatic shut off after 15 minutes (can be disengaged)
PC Interface: USB PC Communication with supplied software and cable for data acquisition
Over range indication: Dashes appears on the LCD
Low battery indication: Battery symbol appears on the LCD

Power supply: 9V Battery
Operating conditions Meter: 0 to 50oC; 80% RH max.
Sensor: 0 to 60oC
Dimensions / Weight Main instrument: 203 x 75 x 50mm
Sensor head: 72mm diameter
Weight: 280g

Ordering Information:
EXT HD300

ext-hd300-spec-sheet

mo-hd300-user-manual

Weight 4 kg
Scroll to Top